Infant Radiant Warmer BNT-2000B
     

Infant Radiant Warmer BNT-2000B

   
 
   
 
Home
Company
Contact
Download
  Products

STANDARD COMPOSITION:


●Temperature Control Mode: pre-heating control, servo control & manual control;
● Instant Parameter Display: set temp, skin temp, heat power rate;
● Built-in APGAR Timer and >37℃ preset function
● Audible/Visual Alarms: temp deviation, sensor failure, power failure, temp override;
● Tilting Bed with Rotary Radiant Head;
● X-ray Cassette Tray under infant bed, examination light of high illuminance;
● Use anti-blast quartz infrared tube as radiant heating source;
● With phototherapy unit;
● With two oxygen cylinders;
● With Suction, with neonate resuscitation.

 

Specifications:

This equipment belongs to ClassⅠ, Type BF, continuous operation common equipment.
POWER
Model BNT-2000-1E:AC 110 / 120 V, 50Hz / 60Hz
Model BNT-2000-2E:AC 220 / 240V, 50Hz / 60Hz
Input Power:≤1000VA
TEMPERATURE CONTROL RANGE AND RELATIVE SPECIFICATION
Control mode::.....Pre-warm mode, manual mode, baby mode
Temperature control range:25℃~37℃,for set temperature ≤37℃
37℃~38℃,for set temperature >37℃
Skin temperature sensor measure range:20℃~42℃
Deviation between sensor measure temperature and control temperature:≤0.5℃
Skin Sensor Precision:≤0.3℃
Bed surface temperature Uniformity:≤1℃
Temperature Rise Time :≤45min
Bed tiltable:continues adjustable in ±10°
ALARM
Sensor failure alarm:alarm with audible and visual when sensor open circuit, short circuit or disconnect, cut off power.
Over temperature alarm:alarm with audible and visual when displayed temperature was approach 38℃ for set temperature ≤37℃,
or approach 39℃ for set temperature >37℃, cut off power.
Deviation alarm:after temperature equilibrium, alarm with audible and visual when displayed temperature was 1℃ above or
below control temperature, cut off power if above 1℃.
Power failure alarm:alarm with audible and visual when power failure
APGAR time:alarm with audible when run between 50 seconds to 1 minute, 4 minutes 50 seconds to 5 minutes, 9 minutes 50 seconds to 10 minutes.
ENVIRONMENT TEMPERATURE
(It is a suggestion that please do not use outside of the allow environment.)
Operate temperature range:18℃~30℃
Store temperature range:-20℃~+55℃
ENVIRONMENT HUMIDITY
Operate humidity range:30%RH~75%RH
Store humidity range:≤93%RH
ATMOSPHERIC PRESSURE
Transport and Store atmospheric pressure range:500hPa~1060hPa
Operate atmospheric pressure range:700hPa~1060hPa
STREAM VELOCITY
Ambit environment stream velocity:≤0.3m/s

   
Ambulance
Infant Incubator
Infant Radiant Warmer
Infant Phototherapy
Jaundice Meter
Patient Monitor
Oximeter
Oxygen Concentrators
Syringe & Infusion Pump
Ultrasound Scanner
ECG Machine
Holter & Stress Test System
Stethoscope
Patient Bed
Ventilator
Ophthalmology
Suction
Endoscope
C-ARM
Anesthetic Machine